AUTOCLAVE 20-25LITERS

SPECIFICATIONS

Type Table top or bench top steam steriliser
Sterilizer chamber capacity in L (usable volume) 20 – 25 liter
Type of cycle processes available B-process,N- process,Dri-Tec drying – process
Sterilizer Chamber Type Rectangular
Steriliser chamber material SS 316 or better grade steel
Working Pressure of Chamber 2.2 bar
Design Pressure of Chamber 3.8 bar
Chamber pre-heating facility: Equipped with electrically heated jacket for preheating on standby mode: No
Type of Sterilizer Chamber Door Hinged Door
Steriliser Chamber door Locking facility At least two locking bolts
Sterilizer Chamber door Operation: Manual
Door sealing: By elastomeric rubber gasket suitable to withstand temperature upto 140 degree C & pressure upto 20-30 psi
Inlet Air Filtration: Yes, Atmospheric air shall enter the sterilisation chamber after its filtering by a disposable air filter provided in the equipment.
Air Filter efficiency: >= 99.998%
Sterilising Cycle Processes available: 121 degree C Wrapped,134 degree C Wrapped,134 degree C Flash/Rapid open instrument cycle,134 degree C textile,Test programs: Bowie & Dick, Leak Test
Sterilizer has water circulation system so that no drain point and fixed water inlets are required: No
Water Storage Tank: Inbuilt water reservoir with Storage capacity of 3-5 litre with easy access for cleaning to avoid biofilm available: Yes
Steriliser have inbuilt steam generator designed for integrated energy storing system for building up power for sterilisation load in shot time: No
Type of control system available for Steriliser cycle operation: Microprocessor based PLC system specifically designed for sterilisation operations, with CPU processor with battery back up, digital input/output controls, analog measuring inputs & COM ports for printer and PC connectivity. Automatic process checking, self checking of safety devices and failure correction should be possible by the control system.
Power Source: 220 Volt, 50 Hz AC electric supply
Power supply: 220 volt, 50 Hz AC electric supply single phase
Number of Trays in sterillser chamber 1
Sterilliser Loading Type: Top Loading
Audio- visual Alarm facility available for notifying: Temperature sensor failure