| Type of Pulse Oximeter |
TABLETOP |
| Should work during motion and very low perfusion conditions |
Yes |
| Special Signal processing Technology (Motion artifact rejection) |
Nellcor Oximax |
| SPO2 probes |
Reusable |
| Oxygen saturation Range |
0-100% |
| Pulse rate |
25-240 BPM |
| Display of SPO2 and PR |
digital display with plethysmograph with trends |
| Accuracy |
± 5 % |
| Type of Patient |
Adult,Pediatric |
| Should be shock proof |
Yes |
| Display (with contrast adjust-ability) |
TFT |
| Display should be Readable from at least 3 feet |
Yes |
| Saturation displayed on the screen |
Yes |
| Pulse Rate displayed on the screen |
Yes |
| Perfusion Index displayed on the screen |
Yes |
| Status of battery charging displayed on the screen |
Yes |
| Sensor off /on status |
Yes |
| Audible and visual Alarms |
Yes |
| Alarm override facility should be present |
Yes |
| Trends |
48 hrs |
| Type of trend display |
Graphical |
| Data interval in sec |
10 |
| Power source |
Inbuilt re-chargeable battery and mains 220-240 V , 50 Hz AC operation |
| Automatic switch from mains to battery in case of power failure |
Yes |
| RS 232C interface for data communication and transfer |
NA |
| Should have provisions for wireless and blue tooth connectivity |
NA |
| Patient extension cable |
Yes |
| Power consumption in watts |
9 |
| Operating Temperature |
“Ambient temperature of 0 to 40 deg C and relative humidity of 15 to 90%” |
| Weight of the machine in kg |
1.5 |
| Protective splash proof case for clean and strogage and safe transport provided |
Yes |
| Length of power cable |
3m or more |
| Noise level in dB |
25 |
| Display must allow easy viewing in all ambient light levels |
Yes |
| Non Invasive Hemoglobin |
No |
| Respiratory Rate (RRP) |
No |
| Pleth Variable Index (PVI) |
No |