Vertical Laminar Air Flow Cabinets
SPECIFICATION
UMDNS Name and Code | Class I |
Other Names | Laminar flow Vertical (LFV) |
Purpose | A laminar airflow workstation is a closed cabinet fitted with HEPA filtered airflow system. Here, laminar means unidirectional constant flow of air with almost no or minimal turbulence. The airflow velocity remains between 0.3 m/s to 0.5 m/s. The purpose of using such workstations in the laboratory is to create a particle and bacteria-free working environment to carry out specialized work. As these units discharge air towards the user, they provide no personal protection but product protection from room contaminants. |
Type of configuration | Freestanding, benchtop |
Type of air flow direction | Vertical |
Type of air cleanliness | Class 100 |
Material used for the work bench | Stainless steel 304 grade |
Type of Sash (front door) | Manual sliding |
Material used for the front door | Acrylic transparent |
Material used for the Side panels | Acrylic |
Thickness of the floor or side panels | 8 millimeter |
Door and Frames should be hermatically sealed | Yes |