LCD Display Orbital Shaking Incubators

SPECIFICATIONS

Construction of Body Double walled with inner and outer walls
Material of Inner wall SS
Material of Outer wall MS with powder coated for anti rusting and duly painted with best finishing
Insulation Proper insulation between inner and outer wall
Insulation material Mineral Wool of 75mm
Door Insulated double walled door with glass window
Handle and casket to be provided Yes
Castor wheels No
Internal length of Equipment in mm 700
Internal width of equipment in mm 700
Internal Height of equipment in mm 600
Outer length of equipment in mm 850
Ouiter width of equipment in mm 950
Outer height of equipment in mm 1500
Warranty in years 3
Temperature Parameters
Range of Temperature 15 C below ambient to 80 C,± 0.1 ? at 37 C in flask
Temperature accuracy plus minus 0.5 degree centigrade
Temperature controller Microprocessor based Digital Temperature controller and indicator
Type of sensor PT -100
Type of Display LCD
Shaking Features
Type of Motor Brushless Motor
Shaking mechanism controller Microprocessor controller
Orbital Shaking speed in RPM 400
Type of speed control Stepless continuously variable
Increment of Step variation 0
Orbital shaking amplitude in mm 25 mm
Timer adjustment in hours 99.99
Availability of Platform to accommodate different types of flasks Yes
Size of platform in mm x mm 18X18
Type of flask which can be accommodated 100 ml,250 ml,500 ml,1000 ml,50 ml,2000 ml,Others,6000 ml
Shaking platform to accommodate test racks Yes
No of clamps for 100 ml flask to be supplied with system 10
No of clamps for250 ml flask to be supplied with system 10
No of clamps for 500 ml flask to be supplied with system 16
No of clamps for 1000 ml flask to be supplied with system 4
Availability of height adjustable and perforated stainless steel shelves above shaking plat form which can accomodate plates and other static samples Yes
Interior illumination LED Bulbs
Digital Display to indicate set RPM Yes
Other Features
Over Temperature Protection Yes
Automatic restart of pre set speed in case of power failure Yes
Shaking to stop when door is opened Yes
Facility to calibrate Temperature and speed Yes
Electronic and mechanical componenets should be enclosed and protected from accidental spilling Yes
Equipment shall be condensation free Yes
Equipment should be provided with proper outlet to drain out liquid sample in case of spillage Yes
Audio visual indication in case of power failure Yes
Audio visual indicator in case of door open Yes
In built lock and key system Yes
Stabilizer provided Yes
Overall weight in Kg 50
Printed product literature to be supplied Yes