Performance Parameters
| Purpose | Dental X ray machine is used for taking Intra Oral Periapical and Occlusal X ray |
| Tube voltage in KV | 65 KV |
| Availability of selection option for Bisecting angle technique & Parallel Technique | Yes |
| Tube current in mA | 2 – 8 mA |
| Focal spot in mm | 0.4 |
| Should have Total filtration and Inherent Filtration | Yes |
| Total filtration in mm Al | 2 |
| Inherent filtration in mm Al | 1.8 |
| Minimum range of exposure time in sec | 0.5 |
| Power supply | 220-240 V , 50 Hz single phase |
| Based on DC Current | Yes |
| RVG / X ray Film compatible | Yes |
| Machine should have a stable base with sturdy locking wheel mechanism | No |
| Machine shall be wall mountable | Yes |
| Should have digital control equipped with an easy ready display indicating with precision | Yes |
| Stool for the patient provided | No |
| Soft positioning arms for accurate tube positions | Yes |
| Light and flexible movements | Yes |
| Internally lead coated head tube and cone to avoid scattered radiations | Yes |
| Exclusive angular indicating system for head positioning in various radiography techniques | Yes |
| High efficiency and greater sharpness of the radiography | Yes |
| Warranty (Option of comprehensive warranty is available through bidding only, which if opted will supersede normal warranty in the catalogue) | 2 |
Additional Parameters
| Should have high voltage generator with high efficiency in the emission of the X rays | Yes |
| Should have free swivel head, which allows easy positioning of head | Yes |
| Number of light weight lead equivalent aprons | 2 (with lead thickness 0.5 mm) |
| To be supplied with Thyroid shield | Yes |
| User manuals to be supplied | Yes |