12 Channels ECG Machine

SPECIFICATION

Purpose ECG machine is primary equipment to record ECG signal with interpretation which is required for recording and analyzing the waveforms with software.
Operating modes of ECG Machine Automatic, Manual and Rhythm
ECG machine should have ECG lead annotation facility Yes
Leads which is in ECG machine should be able to acquire simultaneously and interpret them 12
Number of channels 12
ECG machine should acquire lead ECG for both adult and paediatric patients Yes
The ECG machine should have facility to show lead fail indication Yes
The ECG machine should have facility to show lead reversal indication Yes
The ECG machine should have facility to show the impedance to quality check of connection Yes
Acquisition time for ECG Machine in sec 10 sec
Digital sampling rate for Pacemaker spike detection 40000 s/sec/channel
Recording of digital sampling for pacemaker 1000 s/sec/channel
ECG machine should have real time colour backlit display of ECG waveforms with signal qualify indication for each lead Yes
ECG machine should have frequency filters Artifact, AC and low and high pass frequency filters
Number of ECGs which can be store in ECG Machine 151-200
ECG machine should have full screen preview of ECG report for quality assessment checks prior to print Yes
Type of inbuilt screen LED
Size of screen in inches 7
Display resolution of ECG machine in pixels 640 x 480
ECG machine should have interpretation facility of the amplitudes, duration and morphologies of ECG waveforms and associated rhythm for adult and pediatric patient Yes
ECG machine should have alphanumeric keyboard for patient data entry Hard keys
Size of printing paper A5
Resolution of digital array printer using thermal sensitive paper in ECG Machine in dpi 200 dpi x 500 dpi on 25 mm/sec speed
ECG machine report format Report formats of 3×4; 6×2, Rhythm for up to selected leads, 12 lead extended measurement, 1 minute of continuous waveform data for 1 selected lead
Provision of battery Yes
Battery capacity 50 ECG or 1 hour of continous rhythm recording on single charge
Connectivity to ECG Machine LAN
Storage on external portable memories USB support
The individual patient lead should be change without replacing the whole patient cable assembly Yes
Noise level in dba 25
Weight of ECG Machine in Kg 2.5
Frequency in Hz 300
Availability of latest interpretation software Yes
Wireless acquisition module with RF technology No
System should have the dedicated software to download the ECG form machine in PDF format Yes
Warranty (Option of comprehensive warranty is available through bidding only, which if opted will supersede normal warranty in the catalogue) 1
User and service manual of ECG Machine Yes
Category: