AUTOCLAVE 20-25LITERS
SPECIFICATIONS
Type | Table top or bench top steam steriliser |
Sterilizer chamber capacity in L (usable volume) | 20 – 25 liter |
Type of cycle processes available | B-process,N- process,Dri-Tec drying – process |
Sterilizer Chamber Type | Rectangular |
Steriliser chamber material | SS 316 or better grade steel |
Working Pressure of Chamber | 2.2 bar |
Design Pressure of Chamber | 3.8 bar |
Chamber pre-heating facility: Equipped with electrically heated jacket for preheating on standby mode: | No |
Type of Sterilizer Chamber Door | Hinged Door |
Steriliser Chamber door Locking facility | At least two locking bolts |
Sterilizer Chamber door Operation: | Manual |
Door sealing: | By elastomeric rubber gasket suitable to withstand temperature upto 140 degree C & pressure upto 20-30 psi |
Inlet Air Filtration: | Yes, Atmospheric air shall enter the sterilisation chamber after its filtering by a disposable air filter provided in the equipment. |
Air Filter efficiency: | >= 99.998% |
Sterilising Cycle Processes available: | 121 degree C Wrapped,134 degree C Wrapped,134 degree C Flash/Rapid open instrument cycle,134 degree C textile,Test programs: Bowie & Dick, Leak Test |
Sterilizer has water circulation system so that no drain point and fixed water inlets are required: | No |
Water Storage Tank: Inbuilt water reservoir with Storage capacity of 3-5 litre with easy access for cleaning to avoid biofilm available: | Yes |
Steriliser have inbuilt steam generator designed for integrated energy storing system for building up power for sterilisation load in shot time: | No |
Type of control system available for Steriliser cycle operation: | Microprocessor based PLC system specifically designed for sterilisation operations, with CPU processor with battery back up, digital input/output controls, analog measuring inputs & COM ports for printer and PC connectivity. Automatic process checking, self checking of safety devices and failure correction should be possible by the control system. |
Power Source: | 220 Volt, 50 Hz AC electric supply |
Power supply: | 220 volt, 50 Hz AC electric supply single phase |
Number of Trays in sterillser chamber | 1 |
Sterilliser Loading Type: | Top Loading |
Audio- visual Alarm facility available for notifying: | Temperature sensor failure |