| Purpose |
Used for storage of various samples , sensitive stem cells, plasma , semen , virus, bone graft and other biological samples |
| Capacity in liters |
400-450 L |
| Freezer body |
Freezer should have galvanised steel body with tough powder coated exterior finish constructed on steel |
| Material of inside chamber |
SS |
| Type of Cabinet |
Vertical |
| Ground clearance in mm |
100 mm |
| Should have heavy duty lockable castors and levelling for adjustments and installation |
Yes |
| Refrigerants |
CFC-free, HCFC-free non flammable refrigerants |
| Refrigeration system must be energy efficient with pull down time in hrs |
5.2 hr |
| Operating Temperature with 1ᵒ C increment |
minus 55ᵒC to minus 86ᵒC |
| Freezer must attain -80ᵒC while operating at ambient temperature of 32ᵒC |
Yes |
| Number of holding racks |
15 |
| Control system |
Fully programmable microprocessor controlled with membrane keypad and eye level control panel |
| Number of inner storage compartments with insulated doors and adjustable height |
4 |
| Freezer should have heated air vent and easily accessible front panel air filter |
Yes |
| Digital temperature display |
Yes |
| Should have battery back up for the display and security lock for the display |
NA |
| Alarms |
Alarms for audible and visible alarm system for unwanted temperature rise, sudden power failure , system failure and battery low |
| Noise level in dB |
57 |