Performance Parameters
| Purpose | Autoclaving of equipments, instrument etc for dental treatment |
| Scope of Supply | Auto clave , servo stabilizer,water tank and sterilizatinon pouches |
| Type of Loading | Front |
| Type of auto clave | Table Top |
| No of separators to hold different types of instruments/ materials | 3 |
| Sterilization Process | Automatic |
| Drying process | Automatic |
| Control management and monotoring | Completely microprocessor controlled, constant management and monitoring of autoclave functions and parameters |
| Range of Capacity of autoclave in liters(Hint Seller to select the option based on capacity of auto clave offered) | 20-22 |
| Provision of Pre vaccum cycles to remove air from the chamber | Yes |
| Class type | B |
| Provision of pre vaccum cycles to remove air from the chamber | Yes |
| Provision of Dry cycle option with selectable time periods | Yes |
| Type of display with soft touch pad | LED |
| Material of construction of the perforated trays | Stainless steel 316 L |
| Porosity of microbacterial filter | 0.3 micron |
| Location of Antimicrobial filter | The microbacterail filter is located inside the autoclave undermeath autoclave to prevent dust from going inside the autoclave |
| Weight of the autoclave in KG | 55 |
| Availability of Alpha numeric display to set protocol with soft key pad | Yes |
| Number of error messages and warning systems available | more than 10 |
| Table top auto clave has different pre program cycles of autoclaving different wrapped and unwrapped equipment/materials ie metal solid instruments,hollow metal instruments,rubber ,plastic,textile | Yes |
| Type of monitoring systems available with the Autoclave | Constant monitoring of temperature, pressure, vacuum and drying cycle |
| sterilization holding time for short load | less than or equal to 2 minutes |
| No of preprogrammed autoclaving cycles available | 3 |
| Availability of Prion Program | No |
| sterilization holding time for prion programme in minutes | NA (if Prion program not available) |
| Door opening and closure indication | Acoustic |
| Availability of saftey valve to release excess pressure to prevent accident | Yes |
| Table top autoclave have automatic pressure locking system while sterilization to prevent accidental door opening | Yes |
| Power supply | Supply voltage range 220-240 Volts, 50/60 Hz, single phase |