Ultra low temperature laboratory deep freezer 451-500LITERS

SPECIFICATIONS

Purpose Used for storage of various samples , sensitive stem cells, plasma , semen , virus, bone graft and other biological samples
Capacity in liters 451-500 L
Freezer body Freezer should have galvanised steel body with tough powder coated exterior finish constructed on steel
Material of inside chamber SS
Type of Cabinet Vertical
Ground clearance in mm 100 mm
Should have heavy duty lockable castors and levelling for adjustments and installation Yes
Refrigerants CFC-free, HCFC-free non flammable refrigerants
Refrigeration system must be energy efficient with pull down time in hrs 5.2 hr
Operating Temperature with 1ᵒ C increment 30 degree to -80 degree
Freezer must attain -80ᵒC while operating at ambient temperature of 32ᵒC Yes
Number of holding racks 25
Control system Fully programmable microprocessor controlled with membrane keypad and eye level control panel
Number of inner storage compartments with insulated doors and adjustable height 4
Freezer should have heated air vent and easily accessible front panel air filter Yes
Digital temperature display Yes
Should have battery back up for the display and security lock for the display Yes
Alarms Alarms for audible and visible alarm system for unwanted temperature rise, sudden power failure , system failure and battery low
Noise level in dB 60
Energy consumption rate (@KWh/day) 3.5
Stabilizer should be capable to run any voltage between 190V – 270 V Yes
Double door with locking lids Yes
Availability of decompression valve facility to lower air pressure inside freezer for easy door opening Yes
Panels Vacuum polyurethane foam panels
Number of compressor 1
Temperature uniformity +/- 2 degree
Temperature stability of system ±3ᵒC
Warranty of Stabilizer in Years 1
Warranty of Freezer (except compressor) in Years (from the date of installation) 3
Warranty of compressor in years 1
Category: