Ultra low temperature laboratory deep freezer 95 LITERS
SPECIFICATIONS
| Purpose | NA |
| Capacity in liters | 95 L |
| Freezer body | Freezer should have galvanised steel body with tough powder coated exterior finish constructed on steel |
| Material of inside chamber | SS |
| Type of Cabinet | Vertical |
| Ground clearance in mm | 100 mm |
| Should have heavy duty lockable castors and levelling for adjustments and installation | Yes |
| Refrigerants | Hermetic sealed compressor with CFC free refrigerant |
| Refrigeration system must be energy efficient with pull down time in hrs | less than 5.5 hr |
| Operating Temperature with 1ᵒ C increment | minus 10 degree C to minus 40 degree C |
| Freezer must attain -80ᵒC while operating at ambient temperature of 32ᵒC | Yes |
| Number of holding racks | 15 |
| Control system | Microprocessed Digital Controller system for precise temperature controller |
| Number of inner storage compartments with insulated doors and adjustable height | 1 |
| Freezer should have heated air vent and easily accessible front panel air filter | Yes |
| Digital temperature display | Yes |
| Should have battery back up for the display and security lock for the display | NA |
| Alarms | Electronic alarm if, temperature deviates more than ± 2 C |
| Noise level in dB | 60 |
| Energy consumption rate (@KWh/day) | 24 |
| Stabilizer should be capable to run any voltage between 190V – 270 V | Yes |
| Double door with locking lids | Yes |
| Availability of decompression valve facility to lower air pressure inside freezer for easy door opening | Yes |
| Panels | Vacuum polyurethane foam panels |
| Number of compressor | 2 |
| Temperature uniformity | +/- 3 degree |
| Temperature stability of system | ±3ᵒC |
| Warranty of Stabilizer in Years | 1 |
| Warranty of Freezer (except compressor) in Years (from the date of installation) | 1 |
| Warranty of compressor in years | 1 |